Product characteristics | |||||
---|---|---|---|---|---|
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 | ||||
Measuring range |
| ||||
Process connection | threaded connection G 1/2 external thread | ||||
Installation length EL [mm] | 50 | ||||
Application | |||||
Special feature | Gold-plated contacts | ||||
Measuring element | 1 x Pt 1000 + 1 x NTC; (thermically coupled, with backup function (temperature measuring even if one element fails)) | ||||
Media | liquids and gases | ||||
Pressure rating [bar] | 50 | ||||
Minimum installation depth [mm] | 25 | ||||
MAWP (for applications according to CRN) [bar] | 50 | ||||
Electrical data | |||||
Operating voltage [V] | 18…32 DC; (“supply class 2” to cULus) | ||||
Current consumption [mA] | 6; (24 V) | ||||
Protection class | III | ||||
Reverse polarity protection | yes | ||||
Power-on delay time [s] | 8 | ||||
Integrated watchdog | yes | ||||
Inputs / outputs | |||||
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 | ||||
Outputs | |||||
Total number of outputs | 2 | ||||
Output signal | switching signal; analogue signal; IO-Link; (configurable) | ||||
Electrical design | PNP/NPN | ||||
Number of digital outputs | 1 | ||||
Output function | normally open / normally closed / diagnostic signal; (parameterisable) | ||||
Max. voltage drop switching output DC [V] | 2 | ||||
Permanent current rating of switching output DC [mA] | 150; (200 (…60°C); 250 (…40 °C)) | ||||
Diagnostic output | diagnostic output for drift and fault monitoring | ||||
Number of analogue outputs | 1 | ||||
Analogue current output [mA] | 4…20 | ||||
Max. load [Ω] | (Ub – 15 V) x 50 | ||||
Short-circuit protection | yes | ||||
Type of short-circuit protection | pulsed | ||||
Overload protection | yes | ||||
Measuring/setting range | |||||
Measuring range |
| ||||
Note on measuring range |
| ||||
Factory setting | 0…150 °C / 32…302 °F | ||||
Drift warning |
| ||||
Drift alarm |
| ||||
In steps of |
| ||||
Resolution | |||||
Resolution of analogue output [K] | 0.05 | ||||
Accuracy / deviations | |||||
Precision analogue output [K] | ± 0,2 (-10…130°C); ± 0,3 (130…140°C); ± 0,3 ± 0,1% MS (-25…-10/140…160°C); (probe completely inserted into the measured medium up to the sealing chamfer) | ||||
Temperature coefficient [% of the span / 10 K] | < ± 0,01; (in case of deviation from the reference condition 25 ± 5 °C) | ||||
Response times | |||||
Dynamic response T05 / T09 [s] | 3 / 6 | ||||
Software / programming | |||||
Parameter setting options | drift warning/drift alarm threshold; Fail-Safe; Display unit; scaling of the analogue output; redundancy switching; performance diagnostic output; switching logic; normally open / normally closed | ||||
Interfaces | |||||
Communication interface | IO-Link | ||||
Transmission type | COM1 (4,8 kBaud) | ||||
IO-Link revision | 1.1 | ||||
SDCI standard | IEC 61131-9 CDV | ||||
Profiles | Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis | ||||
SIO mode | yes | ||||
Required master port type | A | ||||
Process data analogue | 1 | ||||
Process data binary | 1 | ||||
Min. process cycle time [ms] | 18.8 | ||||
Supported DeviceIDs |
| ||||
Operating conditions | |||||
Ambient temperature [°C] | -25…70 | ||||
Storage temperature [°C] | -40…85 | ||||
Protection | IP 68; IP 69K | ||||
Mechanical data | |||||
Weight [g] | 335.5 | ||||
Materials | stainless steel (1.4435 / 316L); stainless steel (1.4404 / 316L); PEI; FKM | ||||
Materials (wetted parts) | stainless steel (1.4435 / 316L); stainless steel (1.4404 / 316L) | ||||
Tightening torque [Nm] | 30…50 | ||||
Process connection | threaded connection G 1/2 external thread | ||||
Surface characteristics Ra/Rz of the wetted parts | Ra: < 0.6 | ||||
Installation length EL [mm] | 50 | ||||
Remarks | |||||
Remarks |
| ||||
Pack quantity | 1 pcs. | ||||
Electrical connection | |||||
Connection | Connector: 1 x M12; Contacts: gold-plated |